Agilent Installing the Tray Belt Modification guide

image

Contents

1. Copyright Agilent Technologies 2000 Printed in Germany Edition 02 00 G1600 90109 Part No G1600 90109
2. See Agilent Technologies Installing the Tray Belt Modification G1600 60026 Remove tray base assembly Remove foam parts a and foam insert b from the sample carousel Do not remove the heat exchanger 2 Remove all segments a Loosen screws that hold tray motor b Remove motor tooth wheel by opening the two small hex screws c Remove tray motor and tray belt d Put roller plate on top of foam a Make sure the three holes fit correctly b Use a knife to cut away overlapping foam 4 The new cut out must look like this a Remove all foam cuttings Remaining cuttings inhibit proper function Copyright Agilent Technologies 2000 1 Installing the Tray Belt Modification G1600 60026 5 Insert the roller plate a 6 Put the motor tooth wheel a back and Put the tray motor back b tighten the two hex screws Use the provided screws and fix the motor Insert the bearing sleeve b c Do not tighten it completely because the tension of the tray belt needs to be adjusted 7 Install the longer tray belt a part number 1500 1040 labeled B340 MXL Ensure proper alignment of the tray belt on the belt roller b the tooth wheel c and the tray d Move the motor to give the tray belt a certain tension It is not necessary to over do it Tighten the tray motor screws e