Agilent Dissolution Workstation Installation Setup Instruction

image

Contents

1. NOTICE This document contains references to Varian Please note that Varian Inc is now part of Agilent Technologies For more information go to www agilent com chem Ri Agilent Technologies Dissolution Workstation installation and Setup Instructions P N 70 9049 November 2010 Revision F Limitation of Liability The information in this document is subject to change without notice Varian Inc makes no warranty of any kind with regard to this material including but not limited to the implied warranties or merchantability and fitness for a particular purpose Varian Inc shall not be liable for errors contained herein or for incidental consequential damages in connection with the furnishing performance or use of this material All rights are reserved Reproduction adaptation or translation without prior written permission is prohibited except as allowed under copyright laws First Edition November 2010 Dissolution Workstation Installation and Setup Instructions Part Number 70 9049 Revision F Printed in the United States of America The following terms are trademarks of Varian Inc Benchsaver M Practical Solutions BIO DIS III QA and QAII Enhancer Cell e TruCenter TM Full Flow Filters VanKel Peak Vessel VK Varian Inc 13000 Weston Parkway Cary North Carolina 27513 2250 800 229 1108 919 677 1108 Fax 919 677 1138 www varianinc com VARIAN Copy
2. Revision F 11 10 Dissolution Workstation Page 5 P N 70 9049 Installation and Setup Instructions Installation and Setup Step 3 Click Connect to connect to the server or local PC Step 4 Select the Attach tab on the MSDE Manager and attach the database by clicking Attach Leave all the values at their default settings Step 5 After successful attachment the Database Attach screen displays FIGURE 4 Database Attach screen Database Attach Step 6 Click OK Step 7 Close the MSDE Manager screen Local Security Policy For 21 CFR 11 compliance purposes you should ensure that the following minimum configurations are made to your system s local security policy It may be necessary to coordinate the following configurations with your domain administrator Step 1 Click Start Run to run the Local Security Settings Manager Type secpol msc and press Enter The Local Security Settings Management screen displays Varian Inc Page 6 Dissolution Workstation Revision F 11 10 Installation and Setup Installation and Setup Instructions P N 70 9049 Step 2 Click Security Settings Account Policies Password Policy and set the following security policy configuration Policy Security Setting Enforce Password History 3 passwords remembered Maximum password age 30 days Minimum password length 6 characters Password Must Meet Complexity Requirements Enabled Step 3 Click Security Settings
3. or syringe pump and filter changer Peristaltic Pump Syringe Pump or Syringe Pump and Filter Changer Daisy chain to MDS on page 27 VK 8020 peristaltic pump or syringe pump see HPLC Detector Connection to the VK 8020 on page 14 and Dissolution Apparatus VK 8020 Peristaltic Pump or Syringe Pump Daisy chain to MDS on page 28 VK 8020 syringe pump and filter changer see HPLC Detector Connection to the VK 8020 on page 14 and Dissolution Apparatus VK 8020 Syringe Pump and Filter Changer Daisy chain to MDS on page 29 Varian Inc Page 14 Dissolution Workstation Revision F 11 10 Installation and Setup Installation and Setup Instructions P N 70 9049 HPLC Detector Connection to the VK 8020 If your systems configuration includes a VK 8020 complete the following steps to attach the HPLC Detector to the VK 8020 Step 1 Plug one end of the four pin cable with two bare wires into the outlet labeled Start Output on the rear of the VK 8020 Step 2 Connect the wire labeled 2 to the Start Input negative closure on the HPLC system and the wire labeled 4 to the Start Input positive closure Step 3 Plug one end of the four pin cable with two bare wires into the outlet labeled Injection Input on the rear of the VK 8020 Step 4 Connect the wire labeled 2 to the Ready Out negative closure on the HPLC system and the wire labeled 4 to the Ready Out positi
4. Account Policies Account Lockout Policy and set the following configuration Policy Security Setting Account lockout duration 0 minutes infinite Account lockout threshold 3 invalid login attempts Reset account lockout counter 99999 minutes Step 4 Click Security Settings gt Local Policies gt Audit Policy and set the following configuration Policy Security Setting Audit account logon events Success Failure Audit account management Success Failure Audit logon events Success Failure Audit policy change Success Failure Varian Inc Revision F 11 10 Dissolution Workstation Page 7 P N 70 9049 Installation and Setup Instructions Installation and Setup Configuring the Application Note QE gt Inorder to configure the application it is necessary to be logged on as an administrator After attaching the database complete the following steps to configure the application Step 1 Double click the desktop shortcut to launch the Dissolution Workstation software After the hour glass cursor disappears the application may still take approximately one minute to launch The Configuration Dialog screen displays FIGURE 5 Configuration Dialog screen Configuration Dialog Security Database Miscellaneous User secunty is enabled by adding users to the following user groups Group T Description SES VkDissa uthorizer Allows the user to s
5. P re mS O HORS a 1S Sen Sae P Se ERR P For the appropriate S o NULL Modem jo 5075 0914 cable guide see o e page 17 o 2 o I o Injector o o x DB9 gt DB25 Cable 1 5075 0252 Dissolution E Tester o o e i VK 8020 Peristaltic Pump 2 DB9 gt DB25 Cable 12 0011 5075 0252 Sint RA eed Output Injection Input 5075 044 f Start Input 5075 0085 For the appropriate HPLC Detector 5075 0085 cable c see page 15 a B Ready Out Varian Inc Revision F 11 10 P N 70 9049 Dissolution Workstation Installation and Setup Instructions Page 29 Installation and Setup Dissolution Apparatus VK 8020 Syringe Pump and Filter Changer Daisy chain to MDS L1 HPLC Detector L1 Start Input Dissolution Tester DB9 gt DB25 Cable 5075 0252 Lenos zwooT Le noo lol 5075 0085 fe z Ready Out 4 5075 0085 gt Injection Input Q g VK 8020 Syringe Pump Filter Changer E IN t Start oe Ta m Output 7 jo a lt o ov g s Y Adapter ag o 5075 0446 Sk 2 3 Bo M x DB9 Cable 2 S 5075 0449 T i DB9 DB9 Cable Adapter Y Adapter 9 5075 0449 5075 0446 5075 0446 e Q 1 9 5 5075 0881 o 3 o a E cd a a o e a N O a Q 2 2 a NOTS Oota Cable Guide Part Number Quantity 5075 0446 3 5075 0914
6. and Setup Dissolution Apparatus VK 8000 Syringe Pump Daisy chain to PC VK 8000 Y Adapter 5075 0446 DB9 gt DB25 Cable __ C 5075 0252 1 Y Adapter se 5075 0446 NULL Modem 5075 0914 DB9 gt DB9 Cable 5075 0449 Syringe Pump Cable Guide Part Number Quantity 5075 0446 2 5075 0914 1 5075 0252 2 5075 0449 1 Dissolution Tester DB9 gt DB25 Cable 5075 0252 Varian Inc Page 18 Dissolution Workstation Revision F 11 10 Installation and Setup Installation and Setup Instructions P N 70 9049 Dissolution Apparatus VK 8020 Syringe Pump Daisy chain to PC Start Input E HPLC Detector Leg Ready Qut ai 5075 0085 er 5075 0085 Dissolution Injection Input Syringe Pump Tester Ez Y Adapter al VK 8020 5075 0446 Start m Output T Eq O 2 DB9 gt DB25 Cable __ DB9 DB9 Cable DB9 gt DB25 Cable 5075 0252 j 2 5075 0449 5075 0252 i s um tanto na 5075 0914 Cable Guide Part Number Quantity 5075 0446 2 5075 0252 2 5075 0085 2 5075 0881 1 5075 0449 1 5075 0914 1 Varian Inc Revision F 11 10 Dissolution Workstation Page 19 P N 70 9049 Installation and Setup Instructions Installation and Setup Dissolution Apparatus VK 8000 Syringe Pump Filter Changer Daisy chain to PC E Y Adapter 5075 0446 DB9 gt DB25 Cable l __ DB
7. l l l l l I I l l l l l l l l l l l l l l l l l L Dissolution Tester VK 8000 i Forthe appropriate Dissolution TTS Forthe appropriate cable guide see page 17 For the appropriate cable guide see page 15 z E pue m ee sut e pu Mi pr AUR e um mis ue un oum des nm Gn ani m n m s Tu n m iN ipi cui t um Hub mem ip Ei is pm in 4 9 g VK 8000 Syringe Pump z z 8 o I o C 8 n 8 Y Adapt EI O Y Adapter ag 5075 0446 S5 os 2 52 2 e gt o S N 9 cg ao DB9 gt DB25 Cable 9 8 5075 0252 3 ai DB9 gt DB9 Cable 2 a 5075 0449 g Y Adapter 2 5075 0446 S o NULL Modem 5075 0914 a A EEE E o o 3 J a Peristaltic Pump 12 0011 Varian Inc Page 28 Dissolution Workstation Revision F 11 10 Installation and Setup Installation and Setup Instructions P N 70 9049 Dissolution Apparatus VK 8020 Peristaltic Pump or Syringe Pump Daisy chain to MDS Start Input HPLC Detector mi 5075 0085 Ready Out Injection Input a VK 8020 Syringe Pump goes E 5075 0085 zwo9 i65 z c E L Start LA o al Ei Output d F lQ E 2 o ov g Y Adapter og o e 5075 0446 S5 1L LJ o 1 SQ E c a o S DB9 gt DB25 Cable S g 5075 0252 S DB9 gt DB9 Cable NS S e 5075 0449 Y Adapter amp 5075 0446 a 2 o 9075 088 E J me a Z ial 3
8. 1 5075 0252 2 5075 0449 2 5075 0881 1 5075 0085 2 Varian Inc Page 30 Dissolution Workstation Revision F 11 10 Installation and Setup Installation and Setup Instructions P N 70 9049 This page was intentionally left blank except for this message Varian Inc Index a application configuring 7 attaching the database 4 c cables external interconnecting 2 configuration dialog screen 7 database tab 9 miscellaneous tab 10 configuring the application 7 connecting the detector to the vk 8020 14 d database attach screen 5 database attaching 4 detector to vk 8020 connections 14 e external interconnecting cables 2 i installing the software 2 l license dongle 3 local security policy 5 local users and groups screen 8 logon screen 11 m msde manager screen 4 p pc setting up 1 s setting up your pc 1 software installation 2 v vk 8020 detector connection 14 w workstation connections vk 8000 peristaltic pump daisy chain to mds 27 daisy chain to pc 15 Varian Inc Index individual to mds 21 syringe pump daisy chain to mds 27 daisy chain to pc 17 individual to mds 23 Syringe pump filter changer daisy chain to mds 27 daisy chain to pc 19 individual to mds 25 workstation connections vk 8020 peristaltic pump daisy chain to mds 28 daisy chain to pc 16 individual to mds 22 syringe pump daisy chain to mds 28 daisy chain to pc 18 individual to md
9. 2 D 2 2 o a o T7 Q S co o o Q 2 D 2 2 Varian Inc Page 24 Dissolution Workstation Revision F 11 10 Installation and Setup Installation and Setup Instructions P N 70 9049 Dissolution Apparatus VK 8020 Syringe Pump Individual to MDS DB9 gt DB25 Cable 5075 0252 Start Input Dissolution HPLC Detector Tester DB9 gt DB9 Cable 5075 0448 Ready Out Injection Input Syringe Pump 5075 0085 5075 0085 VK 8020 DB9 gt DB25 Cable 5075 0252 5075 0881 Injector jeuueu 9 9gzo0 seos leuueu2 9 G8Z0 G60S JeAJeg eoieq uod ninWw Cable Guide Part Number Quantity 5075 0252 2 5075 0448 1 5075 0085 2 5075 0881 1 Varian Inc Revision F 11 10 Dissolution Workstation Page 25 P N 70 9049 Installation and Setup Instructions Installation and Setup Dissolution Apparatus VK 8000 Syringe Pump Filter Changer Individual to MDS DB9 gt DB25 Cable 5075 0252 Dissolution Tester DB9 gt DB9 Cable 5075 0448 E Syringe Pump DB9 gt DB9 Cable 5075 0448 B Filter Changer DB9 gt DB25 Cable 5075 0252 VK8000 Cable Guide jeuueu 91 9920 s60G Jeuueu 9 G8Z0 G60S JeAJeg eoweg uod niniw Part Number Quantity 5075 0252 2 5075 0448 2 Varian Inc Page 26 Dissolution Workstation Revision F 11 10 I
10. 9 gt DB9 Cable DB9 gt DB9 Cable DB9 gt DB25 Cable 5075 0252 i 5075 0449 5075 0449 5075 0252 Y Adapter p di 5075 0446 Y Adapter NULL Modem 5075 0446 5075 0914 Cable Guide Part Number Quantity 5075 0914 1 5075 0446 3 5075 0252 2 5075 0449 2 Varian Inc Page 20 Dissolution Workstation Revision F 11 10 Installation and Setup Installation and Setup Instructions P N 70 9049 Dissolution Apparatus VK 8020 Syringe Pump Filter Changer Daisy chain to PC Start Input HPLC Detector 7 Ready Out B 5075 0085 Le 5075 0085 Injection Input Syringe Pump Filter Changer Ed E Y Adapter Vk S020 5075 0446 EE L Start Output 1C dej 2 10 Sy DB9 gt DB25 Cable C DB9 gt DB9 Cabe DB9 gt DB9 Cable DB9 DB25 Cable 3049 0294 gt 5075 0449 5075 0449 5075 0252 5075 0446 ae ES 5075 0881 Y Adapter NULL oe Teon 5075 0446 PC Injector Cable Guide Part Number Quantity 5075 0446 3 5075 0914 1 5075 0252 2 5075 0449 2 5075 0881 1 5075 0085 2 Varian Inc Revision F 11 10 Dissolution Workstation P N 70 9049 Installation and Setup Instructions Page 21 Installation and Setup Dissolution Apparatus VK 8000 Peristaltic Pump Individual to MDS DB9 gt DB25 Cable 5075 0252 Dissolution Tester DB9 gt DB25 Cable 5075 0252 VK 8000 12 0011 Peristaltic Pump Cable Guide P
11. 9049 Installation and Setup Instructions Installation and Setup Step 7 Click OK The Logon screen displays see Figure 9 Logon screen on page 11 The application has been configured FIGURE 9 Logon screen Dissolution Workstation VARIANNC This software is intended for use by authorized personnel only t cm Workstation Connections There are three ways to connect your dissolution equipment to the Dissolution Workstation See the table on page 12 Each dissolution system requires at least one serial port If the system includes the injector module a second serial port is required A typical computer has two serial ports and therefore has the capacity to connect up to two dissolution systems if the injector module is not used or one system if the injector is used If your system configuration includes a VK 8020 see HPLC Detector Connection to the VK 8020 on page 14 in addition to the appropriate connection diagram Cable guides are listed with each connection diagram to aid in selecting the necessary cables for your configuration Since the software can control up to four complete dissolution systems it is often necessary to add additional serial ports to the PC controlling the instruments We recommend the use of either the Edgeport 8 8 serial DB 9 or Edgeport 416 4 USB 16 serial DB 9 Varian Inc Page 12 Installation and Setup Dissolution Workstation Installation and Setup Instructions R
12. art Number Quantity 5075 0252 2 12 0011 0 cz Ei o A Oo o o o n D 3 o S a a o o q o N ez e Q gt D 3 5 o a o o T o bs o o Q gt D 5 5 o supplied with pump Varian Inc Page 22 Dissolution Workstation Revision F 11 10 Installation and Setup Installation and Setup Instructions P N 70 9049 Dissolution Apparatus VK 8020 Peristaltic Pump DB9 gt DB25 Cable Ready Out 5075 0252 DB9 gt DB25 Cable 5075 0252 Start Input Dissolution Tester Injection I njection Input 5075 0085 lt 5075 0085 5075 0881 Start Output Peristaltic Pump jguueu 9L 98Z0 S60S Jeuueu 8 S920 S609 Janes eoi eg uod niniWw Cable Guide Part Number Quantity 5075 0252 2 5075 0881 1 5075 0085 2 12 0011 0 supplied with pump Varian Inc Revision F 11 10 Dissolution Workstation Page 23 P N 70 9049 Installation and Setup Instructions Installation and Setup Dissolution Apparatus VK 8000 Syringe Pump Individual to MDS DB9 gt DB25 Cable 5075 0252 Dissolution Tester DB9 gt DB25 Cable 5075 0252 a VK 8000 DB9 gt DB9 Cable 5075 0448 8 Syringe Pump Cable Guide Part Number Quantity 5075 0252 2 5075 0448 1 i zm o o es O D lt o n D lt o E gt a o T o N co e Q
13. evision F 11 10 P N 70 9049 Note When using the multiport device server you can connect different dissolution systems based on port availability This is true with daisy chain or individual connection See Dissolution Apparatus VK 8000 Peristaltic Pump Syringe Pump or Syringe Pump and Filter Changer Daisy chain to MDS on page 27 The following instructions refer to COM1 COM2 etc It is acceptable to substitute different COM port numbers based on the physical layout of your system Workstation Connection Daisy chain Connection to PC Peripheral Equipment Procedures VK 8000 peristaltic see Dissolution Apparatus VK 8000 pump Peristaltic Pump Daisy chain to PC on page 15 VK 8020 peristaltic see HPLC Detector Connection to the VK 8020 pump on page 14 and Dissolution Apparatus VK 8020 Peristaltic Pump Daisy chain to PC on page 16 VK 8000 syringe see Dissolution Apparatus VK 8000 Syringe pump Pump Daisy chain to PC on page 17 VK 8020 syringe see HPLC Detector Connection to the VK 8020 pump on page 14 and Dissolution Apparatus VK 8020 Syringe Pump Daisy chain to PC on page 18 VK 8000 syringe see Dissolution Apparatus VK 8000 Syringe pump filter changer Pump Filter Changer Daisy chain to PC on page 19 VK 8020 syringe pump filter changer see HPLC Detector Connection to the VK 8020 on page 14 and Dissolutio
14. ign final acceptance of the results BE VkDissoRunMethods Allows the user to run any method on any system Si ViDissoDiagnostics Allows the user to change system configuration settings or to add nev BE VkDissoSystemConhg Allows the user to run diagnostic utilities on a configured system BE ViDissoMethodMaint Allows access to Method Creation snd modification functions BE VkModfyOthercMethod Allows modification of methods owned by other users BE VkModfyOthersSystems Allows modification of systems owned by other users Is Any user that is not member of one of these groups wil have minimal access to reporting functions through membership of the Everyone group User Administration x ie Varian Inc Page 8 Dissolution Workstation Revision F 11 10 Installation and Setup Installation and Setup Instructions P N 70 9049 Step 2 To add a user to a group select the Security tab on the Configuration Dialog screen Note IE This screen is also accessible within the Dissolution Workstation software by clicking Tools and selecting Options from the menu bar Only an administrator can add users Click User Administration at the bottom of the screen The Local Users and Groups screen displays FIGURE 6 Local Users and Groups screen Local Users and Groups fie Acton View Hep LE E xcv G3 EQ Local Users and Groups Local Hame Description La CJ users Power users Power Users posses
15. issolution equipment see Workstation Connections on page 11 Software Installation Note The CD contains all of the requirements for an installation To complete the installation of the Dissolution Workstation software you must logon to the PC as an administrator To install the Dissolution Workstation software complete the following steps Note IE If you are upgrading the software do not uninstall the previous version Execute setup exe to upgrade your installation Step 1 Execute the file setup exe and follow the on screen prompts Step 2 When asked if you want to install the Microsoft Net Framework you must click Yes Varian Inc Revision F 11 10 Dissolution Workstation Page 3 P N 70 9049 Installation and Setup Instructions Installation and Setup Step 3 After the installation has successfully completed install the license dongle by plugging it into an open USB port on your computer The Windows operating system automatically begins communication with the security dongle and allows you to execute the Dissolution Workstation application FIGURE 1 License dongle Step 4 In Windows click Start gt Control Panel gt Power Options Under Power schemes select Always On FIGURE 2 Power Scheme Power Options Properties Power Schemes Alams Power Meter Advanced Hibemate Select the power scheme with the most appropriate settings for this computer Note
16. n Apparatus VK 8020 Syringe Pump Filter Changer Daisy chain to PC on page 20 Varian Inc Revision F 11 10 P N 70 9049 Dissolution Workstation Installation and Setup Instructions Page 13 Installation and Setup Workstation Peripheral Connection Equipment Procedures Individual VK 8000 peristaltic see Dissolution Apparatus VK 8000 Connection to pump Peristaltic Pump Individual to MDS on page 21 Multiport Device VK 8020 peristaltic see HPLC Detector Connection to the VK 8020 Server MDS pump on page 14 and Dissolution Apparatus VK 8020 Peristaltic Pump Individual to MDS on page 22 VK 8000 syringe see Dissolution Apparatus VK 8000 Syringe pump Pump Individual to MDS on page 23 VK 8020 syringe see HPLC Detector Connection to the VK 8020 pump on page 14 and Dissolution Apparatus VK 8020 Syringe Pump Individual to MDS on page 24 VK 8000 syringe see Dissolution Apparatus VK 8000 Syringe pump filter changer Pump Filter Changer Individual to MDS on page 25 VK 8020 syringe see HPLC Detector Connection to the VK 8020 pump filter changer on page 14 and Dissolution Apparatus VK 8020 Syringe Pump Filter Changer Individual to MDS on page 26 Daisy chain VK 8000 peristaltic see Dissolution Apparatus VK 8000 Connection to Multiport Device Server MDS pump syringe pump
17. nstallation and Setup Installation and Setup Instructions P N 70 9049 Dissolution Apparatus VK 8020 Syringe Pump Filter Changer Individual to MDS DB9 gt DB25 Cable 5075 0252 Start Input Dissolution HPLC Detector Tester DB9 gt DB9 Cable mm 5075 0448 a Syringe Pump Injection Input 5075 0085 5075 0085 VK 8020 DB9 gt DB9 Cable 5075 0448 Filter Changer Output DB9 gt DB25 Cable 5075 0252 5075 0881 Injector jeuUeYD 9 99z0 seos leuueu 9 S8Z0 G60S JeAJeg eoweq uod niniw Cable Guide Part Number Quantity 5075 0252 2 5075 0448 2 5075 0881 1 5075 0085 2 Varian Inc Revision F 11 10 P N 70 9049 Dissolution Workstation Installation and Setup Instructions Page 27 Installation and Setup Dissolution Apparatus VK 8000 Peristaltic Pump Syringe Pump or Syringe Pump and Filter Changer Daisy chain to MDS NULL Modem 5075 0914 DB9 DB25 Cable 8 NULL Modem i 5075 0914 5075 0252 DB9 gt DB25 Cable 5075 0252 Y Adapter 5075 0446 5 L L L 2 o o l oe ov Y Adapter ag 5075 0446 Sk 1L 2 1L 2 SQ 1 DB9 gt DB9 Cable DB9 gt DB25 Cable E 5075 0449 5075 0252 E MM DB9 gt DB9 Cable 5075 0449 2 Y Adapter Y Adapter 5075 0446 5075 0446 c WOO l l l l l l l l l l l l l I l l I
18. right c 2005 2010 by Varian Inc Table of Contents Installation and Setup 1 Setting up Your PC 1 External Interconnecting Cables 2 Software Installation 2 Attaching the Database 4 Local Security Policy 5 Configuring the Application 7 Workstation Connections 11 HPLC Detector Connection to the VK 8020 14 Dissolution Apparatus VK 8000 Peristaltic Pump Daisy chain to PC 15 Dissolution Apparatus VK 8020 Peristaltic Pump Daisy chain to PC 16 Dissolution Apparatus VK 8000 Syringe Pump Daisy chain to PC 17 Dissolution Apparatus VK 8020 Syringe Pump Daisy chain to PC 18 Varian Inc Table of Contents Dissolution Apparatus VK 8000 Syringe Pump Filter Changer Daisy chain to PC 19 Dissolution Apparatus VK 8020 Syringe Pump Filter Changer Daisy chain to PC 20 Dissolution Apparatus VK 8000 Peristaltic Pump Individual to MDS 21 Dissolution Apparatus VK 8020 Peristaltic Pump Individual to MDS 22 Dissolution Apparatus VK 8000 Syringe Pump Individual to MDS 23 Dissolution Apparatus VK 8020 Syringe Pump Individual to MDS 24 Dissolution Apparatus VK 8000 Syringe Pump Filter Changer Individual to MDS 25 Dissolution Apparatus VK 8020 Syringe Pump Filter Changer Individual to MDS 26 Dissolution Apparatus VK 8000 Peristaltic Pump Syringe Pump or Syringe Pump and Filter Changer Daisy chain to MDS 27 Dissolution Apparatus VK 8020 Pe
19. ristaltic Pump or Syringe Pump Daisy chain to MDS 28 Dissolution Apparatus VK 8020 Syringe Pump and Filter Changer Daisy chain to MDS 29 Varian Inc Installation and Setup Setting up Your PC The Dissolution Workstation software requires Microsoft Windows 2000 or Windows XP to be installed on your computer For instructions on installing the operating system refer to the documentation supplied with Windows Minimum PC Configuration Windows 2000 SP6 or 256 MB RAM Windows XP SP1 1024 x 768 pixel display resolution 10 GB free hard disk drive space 2 USB ports or a USN port and an RS232 port If the PC is not equipped with a standard serial port a USB to serial converter must be purchased Varian supports the use of the Edgeport USB serial converters manufactured by DIGI International If Varian is installing the Dissolution Workstation software for you note that the installation of a Windows operating system is not included as part of the standard instrument installation Varian Inc Page 2 Dissolution Workstation Revision F 11 10 Installation and Setup Installation and Setup Instructions P N 70 9049 External Interconnecting Cables The monitor keyboard mouse and printer plotter are connected to the PC via cables which plug into the back of the PC Consult your monitor and printer manuals for details of their individual cabling requirements To make cabling connections for the d
20. s 24 syringe pump filter changer daisy chain to mds 29 daisy chain to pc 20 individual to mds 26 Varian Inc
21. s most administr I Gaps Remote Desktop Users Members inthis group aro granted t Sa Roplcator Supports file replication in a doman Users Users aro prevented from making ac ChemStation onagers CheenStynon Managers Group ChemStationOperators Chenctation Operators Group Debugger Users Debugger users can debug processe HelpServicesGroup Greup for the Help and Support Center E SheredChemManacers Shared ChemSiation Manager Logins SheredChemOperators Shared ChemSation Operator Logins HiVidicsoAuthorizer Allows the user to sign final accepts I i B vicisecDingnostics Alas the user to change system co G viciscottetheatent Alas access to Method Cresten an Bviciscohurttethods Alias the user to run any methed o B viciccoSystemContig Allows the user to run deqnostic util GBvittedtyCthersttethed _Alloavs modification of methods ovme GvitrodtyCehersSystems Allows modification of systems oana ius Cevelopers Visual Studio developers can author a Double click the desired VKxxxx group The Group Property screen displays Click Add to add users to the group The Select Users Computers or Groups screen displays Enter the names of the users to add to the group and click OK Repeat step 2 as appropriate for each group that requires new users Varian Inc Revision F 11 10 Dissolution Workstation Page 9 P N 70 9049 Installation and Setup Instructions Installa
22. that changing the settings below will modfy the selected scheme Settings for Aways On power scheme When computer is Ye Pei cue on Tum off monitor v After 15 mins Tum off hard disks v After 30 mins System standby Never System hibemates Varian Inc Revision F 11 10 Dissolution Workstation P N 70 9049 Page 4 Installation and Setup Instructions Installation and Setup Attaching the Database Once the software installation is complete it is necessary to link the Dissolution Workstation software to the database that is installed on the PC as part of the software installation To attach the database locally complete the following steps Step 1 Click Start gt All Programs gt Varian gt Dissolution gt MSDE Manager The MSDE Manager screen displays FIGURE 3 MSDE Manager screen ZB MSDE Manager Utities Servers Available Servers DEONLABNVARIANDISSO p Use Trusted Connecbon Backup Attach MDF File of database to attach C Program Files Microsoft SQL Server MSSOLSVARIANDISSO Data Dissowkstn mat Attach As Dism Attach Step 2 Select a machine name from the list of available servers When connecting to a database installation on a local machine the encrypted password is automatically entered in the password box When connecting to a remote database your system administrator may have to provide you with a password Varian Inc
23. tion and Setup Step 3 From the Configuration Dialog screen select the Database tab and select the SQL server that hosts the database FIGURE 7 Database tab Configuration Dialog Security Database Miscellaneous Database Server VARIANDISSO VARIANDISSO Test Connection Note If your local server is not listed the service may not be running qr Enter services msc Press Enter If necessary start the MSSQL VARIANDISSO service Otherwise manually enter the database server as ComputerName VARIANDISSO where ComputerName is the name of your computer Step 4 Click Test Connection to ensure the database connection is functional Step 5 Select the Miscellaneous tab To run the software without physically Check the status in the Services applet by clicking on Start Run connecting instruments check the Use Simulated Instruments box The system operates in simulation mode Varian Inc Page 10 Dissolution Workstation Revision F 11 10 Installation and Setup Installation and Setup Instructions P N 70 9049 Step 6 Enter an appropriate length of time in minutes the application can be idle before locking FIGURE 8 Miscellaneous tab Configuration Dialog Use Simulated Instruments Number of idle minutes before the application is automaticaly locked 15 Minutes Varian Inc Revision F 11 10 Dissolution Workstation Page 11 P N 70
24. ve closure Varian Inc Revision F 11 10 Dissolution Workstation Page 15 P N 70 9049 Installation and Setup Instructions Installation and Setup Dissolution Apparatus VK 8000 Peristaltic Pump Daisy chain to PC Pump Control s VK 8000 Fraction Dissolution Peristaltic Pump 12 0011 Collector Tester DB9 gt DB25 Cable DB9 gt DB25 Cable 5075 0252 5075 0252 Y Adapter 5075 0446 Ec ii COM 1 COM2 NULL Modem 5075 0914 PC Cable Guide Part Number Quantity 5075 0446 1 5075 0914 1 5075 0252 2 12 0011 0 supplied with pump Varian Inc Page 16 Dissolution Workstation Revision F 11 10 Installation and Setup Installation and Setup Instructions P N 70 9049 Dissolution Apparatus VK 8020 Peristaltic Pump Daisy chain to PC Start Input f HPLC Detector Ready Out 5075 0085 a 5075 0085 Pump Control ZU Injection Input Y A Peristaltic Pump VK 8020 12 0011 Dissolution Tester X Start Output DB9 gt DB25 Cable DB9 gt DB25 Cable 5075 0252 gt 5075 0252 Y Adapter 5075 0446 5075 0881 NULL Modem 5075 0914 PC Injector Cable Guide Part Number Quantity 5075 0085 2 5075 0252 2 5075 0446 1 1 1 0 5075 0914 5075 0881 12 0011 supplied with pump Varian Inc Revision F 11 10 Dissolution Workstation P N 70 9049 Installation and Setup Instructions Page 17 Installation