Agilent Technologies KMW200B User Manual

image

Contents

1. off on mode ro ext temperature real mode ro unit 0 1C ext temperature limit real mode rw unit 0 1C ext temperature state ok warn error mode ro ext temperature valve open real mode ro unit 0 1C ext valve off on mode rw help hwid string mode ro init integer mode rw int flow real mode ro unit 0 01 min int flow const integer mode ro range 200 10000 int flow min real mode ro unit 0 01 min int flow state ok error mode ro int name string mode rw int power off on mode rw int pump off on mode rw int refill off on mode rw int temperature real mode ro unit 0 1C int temperature limit real mode rw unit 0 1C int temperature state ok warn error mode ro int temperature target real mode rw unit 0 1C int water level ok low mode ro state ok error mode ro status integer mode ro store integer mode ro Below is a list of how to use some of the commands auto on module automatically reports new parameter values cooling power returns the present cooling power int temperature returns present internal circuit temperature int flow returns present internal circuit flow rate int flow min returns sets the error limit for flow rate int temperature target int name int temperature limit ext temperature ext temperature valve open ext temperature limit int refill on
2. To do this first remove the front grill panel at the bottom of the rack by unscrewing two screws Push water hoses and electrical cables under the rack and out of the front side so that they can be attached to the rear panel of the KMW200B Roll the KMW200B under the rack and secure in place with two screws The bracket holding the small water filter should be mounted on the side of the electrical rack towards the rear The water cooler has a separate control panel with mains power switch and indicator lamps This panel should be mounted under the other rack mounted devices in the electrical rack 5 2 1 3 Electrical Services The electrical supply must conform to the requirements stated in the SPECIFICATIONS section of this manual Locate the electricity supply close to the system and provide Phase L Neutral N Protection PE The mains plug should be readily reachable by the operator when the equipment has been installed 5 2 1 4 Water Supply Mount the supplied water filter on the main water supply pipe before the connection to the KMW200B in order to prevent any contaminating particles from entering the water cooler The state of the filter cartridge should be checked at regular intervals and replaced when it is necessary 5 2 2 Setting to Work The location of connectors on the KMW200B is shown in figure 5 2 1 Figure 5 2 1 Electrical and Water Connections on the Back of the KMW200B KMW200B Version 15 USER MANUAL Page 10 HAND
3. safe lifting procedures for heavy items to prevent possible strain injury 1 4 Safe Mechanical Practice In normal use personnel are not required to undertake mechanical work However servicing or repair may necessitate access to any part of the system Only suitably qualified personnel should attempt to dismantle modify or repair equipment Water connections should be made and tested in accordance with any local and national safety regulations 1 5 Modifications and Service The manufacturer will not be held responsible for the safety reliability or performance of the equipment unless assembly operations extensions re adjustments modifications and repairs are carried out only by persons authorised by the manufacturer It should be stressed that those parts of the equipment which are interchangeable and which are subject to deterioration during operation may significantly affect the safety of the equipment KMW200B Version 15 USER MANUAL Page 3 INTRODUCTION 2 Introduction 2 1 Scope This manual applies to the KMW200B water cooler unit designed and manufactured by Agilent Technologies 2 2 How to Use This Manual This manual is aimed at operators of the KMW200B who should be trained laboratory technicians and should have been trained to use the KMW200B by Agilent Technologies personnel This manual is intended to provide operators with a practical guide to the system and its operation This is intended to familiaris
4. store store 2 status KMW200B USER MANUAL returns sets the target for the internal circuit temperature returns sets the name of the module Ccd returns sets the error limit for internal circuit temperature returns present external circuit temperature returns the temperature at which the external circuit valve is opened returns sets the error limit for external circuit temperature manually opens refill valve for 3 seconds returns cumulative number of stored parameter changes value 2 is the password to store changes and cumulative number of changes is returned returns status information Version 1 5 Page 15 MAINTENANCE SCHEDULE 7 Maintenance Schedule 7 1 Introduction Maintenance must be performed to ensure that the KMW200B system continues to operate safely and reliably This is detailed in the maintenance schedule given below WARNING Failure to perform scheduled maintenance tasks properly and at the correct intervals can affect the safety and performance of this system AN WARNING Before performing any maintenance task ensure that you have read and understood the HEALTH AND SAFETY INFORMATION at the beginning of this manual Planned maintenance that can be performed by the user is limited to cleaning the water filters Other tasks should be carried out by the user s authorised service representative 7 2 Annual Maintenance Schedule This schedule may have to be carried out more frequent
5. unit has shut down 6 5 Remote control by Ethernet The unit has an Ethernet connection The parameters of the connection can be modified using a web browser with Java This will allow the port number and IP address to be modified By using a program such a hyperterminal or a telnet connection communication can be started with the water cooler module A TCP IP connection can be opened by using the IP address and port number Default port number is 2500 Default IP addresses are 192 168 254 4 CCD cooler 192 168 254 1 Connected PC Text commands can be sent to the module and text replies will be sent back Any changes made to parameters will not be stored after powering down the unit unless the command store 2 is sent the value 2 is treated as a password to activate permanent saving of the parameters The command will send a reply indicating the accumulated number of changes that have been stored in the module Pressing ENTER to send a blank line will result in the unit replying with its name for example IS cooler KMW200 1 1 5 103 Entering help will result in the unit replying with a list of its commands In this list the character is used to represent a space character An example of the list is below the actual list may be different due to product development KMW200B Version 15 USER MANUAL Page 14 OPERATING INSTRUCTIONS auto off on mode rw cooling power integer mode rw range 0 100 cooling state
6. LING INSTALLATION STORAGE AND TRANSIT INFORMATION 5 2 2 1 Equipment Required Power cable for Agilent Technologies CCD detector Mains power cable for KMW200B Water pipes thickness 10 x 3 mm Panel containing water cooler power switch and indicator lamps Signal cable to drive indicator lamps Ethernet cable for communication between water cooler and system control PC Internal circuit small water filter mounted on metal bracket External circuit large water filter and wall mounting bracket 5 2 2 2 Setting up Procedures 1 Connect the water pipes between the CCD detector and the internal circuit of the KMW200B including the small water filter on the return hose Refer to figure 5 2 2 NOTE Water can flow through the Agilent Technologies CCD detector in either direction E Sapphire Internal circuit small filter Pipes 10 3mm lt Y KMW200B Figure 5 2 2 Water pipe connection between KMW200B and CCD detectors Connect the mains water supply after the large water filter to the external circuit inlet of the water cooler Connect the external circuit outlet to the mains water return point e g drain or in house water circuit return pipe The drain connection on the rear panel must be left open to atmospheric pressure Small amounts of water may be ejected from this air vent so a hose should be connected to an open sink drain Connect the indicator lamps cable from the separate front panel to the soc
7. Version 1 5 September 2010 Agilent Technologies XRD Products 10 Mead Road Yarnton Oxfordshire OX5 10U UK Tel 44 0 1865 291600 Fax 44 0 1865 291601 http www agilent com chem KMW200B Water Chiller for CCD detectors User Manual ie Agilent Technologies Important Information This user manual is intended to help the user operate the KMW200B water cooler unit manufactured in Poland by Agilent Technologies Product KMW200B Model Type CCD cooler unit Electrical Ratings 1 N AC 90 264 V 50 60 Hz 200 Watts Before attempting to operate the system PLEASE READ THE INSTRUCTIONS This product should only be used by persons legally permitted to do so If the equipment is used in a manner not specified in the User Manual the protection provided by the equipment may be impaired Important Health and Safety Notice When returning components for service or repair it is essential that the item is shipped together with a signed declaration that the product has not been exposed to any hazardous contamination or that appropriate decontamination procedures have been carried out so that the product is safe to handle Care has been taken to ensure the information in this manual is accurate and at an appropriate level Please inform Agilent Technologies if you have any suggestions for corrections of improvements to this manual This users manual has been written according to standard 89 392 EEC and further modifica
8. ace the bucket under the used filter housing to catch any water spillage Using the specialised tool unscrew the filter housing Pour water out of the housing and remove the old cartridge St 2 3 1 Place the new cartridge into the housing and refit the housing using the tool KMW200B Version 15 USER MANUAL Page 18 MAINTENANCE INSTRUCTIONS Directional valve Pressure reducer Inlet pipe Outlet pipe N Use specialised tool to remove this locking nut Filter cartridge inside housing Figure 8 2 1 Removing the Water Filter Cartridge KMW200B Version 15 USER MANUAL Page 19 9 Trouble Shooting TROUBLE SHOOTING Problem Cause Action Unit cannot be turned on from the front panel Mains power is not connected Connect mains power at the rear panel and at the mains distributor in the system rack Level error LED flashes after switching on module Water level is too low in the closed circuit Water cooler needs time to automatically fill water into the closed circuit especially when starting from a dry system Wait one minute to see if LED stops flashing and turn the module off and on again Flow error LED flashes during operation Water cannot flow through the internal circuit through CCD detector Check pipes are properly connected and valves are open Check filter is not blocked and clean if necessary Water supply error LED flashes duri
9. al Services Power connection Maximum power consumption Maximum mains current CCD output power consumption Main fuse 1 N AC 90 264V 50 60 Hz 200 W 25A Up to 250 W 13 15 A 3 4 External Water Supply Pressure Temperature stability Temperature range Composition Connections 0 5 3 bar gauge 5C 10 20 C Filtered without deposits chemically neutral and optically clear Designed for pipes of size 10 x 3 mm secured with hose clamps KMW200B USER MANUAL Version 1 5 Page 5 TECHNICAL DESCRIPTION 4 Technical Description 4 1 Overview Figure 4 1 1 Front View of KMW200B Water Cooler The KMW200B Cooler consists of a small water reservoir cooling system flow pump and control circuitry It is located under the system electrical rack close to the floor The power switch and indicators are located on a separate front panel at the bottom of the electrical rack They are accessed by opening the front door of the electrical rack Electrical and water connections are located on the back of the cooler They are accessed by unscrewing the front panel of the cooler and rolling it out from underneath the electrical rack Mia mm HERE Figure 4 1 2 Back View of the KMW200B Water Cooler KMW200B Version 15 USER MANUAL Page 6 TECHNICAL DESCRIPTION Figure 4 1 3 Separate front panel of the KMW200B Water Cooler 4 2 Operating Principles The Water Cooler is connected
10. amage to the equipment 1 Never connect disconnect the CCD detector power cable from the KMW200B when the unit is turned on 2 Ensure that the earth cable is properly connected to the KMW200B casing 6 2 Controls and Indicators The controls and indicators of the KMW200B are located on the separate front panel The KMW200B is switched on off using the mains switch on the separate front panel There are four error indicator lights marked Flow Level Temp and Water Supply The four lights are illuminated during the automatic water conditioning check which lasts about 3 seconds 6 2 1 Error status LEDs The error states shown by the LEDs are Flow Level Temperature and Water Supply No LEDs shine during normal operation The Flow error LED shines if water cannot flow in the internal circuit The Level error LED shines if there is insufficient water in the internal circuit and this cannot be filled from the mains water supply The Temperature LED shines if the temperature of the internal circuit water is outside of normal limits The Water Supply LED shines if the temperature of the external circuit water is outside of normal limits and shows that either the mains water is not flowing quickly enough or it is too hot 6 3 Switch on Procedure 1 Check that the water pipes and electrical cables are properly connected 2 Switch on the KMW200B by using the switch on the front panel The lights on the fro
11. ase from the bottom AN CAUTION Do not remove the equipment from the packing crate until it has been moved to its designated installation site The equipment has been carefully packed to protect the equipment from damage in transit Removal of the packing equipment could make the equipment vulnerable to damage during transit 4 Always lift the packing case from the bottom using suitable lifting equipment refer to list of component weights in the following section 5 Move the packing case into the designated installation site 5 1 2 Unpacking 1 Retain all packing material until installation of the system is completed 5 1 3 Weights Dimensions and Lifting Points Description Weight Dimensions Centre of Lifting points kg width x height x depth gravity cm KMW200B Cooler 19 60 x 10 x 37 Right hand Underneath at the side of unit four corners KMW200B Version 1 5 USER MANUAL Page 9 HANDLING INSTALLATION STORAGE AND TRANSIT INFORMATION 5 2 Installation and Setting to Work 5 2 1 Preparation of Site and Services 5 2 1 1 Environmental Requirements It is the customer s responsibility to ensure that all local building and safety regulations are met Ensure that the environmental conditions of the installation site conform to the requirements stated in the SPECIFICATIONS section of this manual 5 2 1 2 System Layout Place the KMW200B water cooler underneath the electrical rack of the system
12. cccccscsscssssessssessssesesesvesesvcsesscsesvssssvcsesscsssvcsssvssssccsssccsssccsssccsssccassccsssccisseesssessssecassceassceassceeeaeess 8 4 3 2 Electrical CONO Sn LR eh Re 8 5 Handling Installation Storage and Transit Information 9 5 1 Reception and Handling coi aa ii dial oia 9 E BY AAA A A A E A E AN SES ORES 9 A hess NE header heise dashed ieee 9 5 1 3 Weights Dimensions and Lifting Points c ccccscsssssessessessesssssssssessessessesssssssussessesseeseessassaucstesesseeseeseaneans 9 5 2 Installation and Setting to Work gxis a ieii 10 5 21 Preparation Of Site and EVE iaa 10 5 2 1 1 Environmental Requirement cccccscscssescssescssesessesecsesccsesecsesecsssccsesccssscesssccssseeasseeassceesscensaceasseess 10 5 2 1 2 System Layout 5 2 1 3 Electrical Services ANN dd Se es ERE NS 5 2 2 Setting tO Work scesascascetesesacsdecasasscessassniessasdasoscacegeactacsdesaissatesadedatsasaadadeassadadeassaisdesatsdafescsedadsasdndadsantachsecnsadedss 52 2 t A esisaoveseadi sivas ere eres ere ra RE 11 52 22 Setting Up Procedures cial 11 A teo E ER DERE RES 12 6 Operating Instructions nd 13 6 1 Warnings and Cautions ii a ad aa 13 6 2 Controls and Indicators cccesssssessssesssssessessesssssssscsssssssessessessesssasssussussusssssessessssusansaussussesseeseessansaucaussesseeaseseeneaneanss 13 6 21 Error status LED e leo te 13 KMW200B Version 15 USER MANUAL Page iii CONTENTS 6 3 SWItCH ON ProcedUl iiewitvait
13. e the operator with how the system works and provide a better understanding of the system operation All personnel who are likely to operate the system or come into contact with any of the system components should read the SAFETY section of the manual This provides basic information aimed at highlighting the safety hazards associated with the equipment The purpose of this manual is to e explain how to operate the equipment e explain how to interface to the equipment e list performance characteristics of the equipment e describe how the equipment operates e assist with simple fault finding and maintenance 2 3 System Description The KMW200B is a stand alone cooler unit intended for indoor use It is designed for use with Agilent Technologies CCD detectors The cooler has the following purposes To stabilise the temperature of the cooling water that is supplied to the CCD detector To electronically control the flow of water to the CCD detector To control the supply of power to the CCD detector KMW200B Version 15 USER MANUAL Page 4 3 Specifications 3 1 Environment SPECIFICATIONS Ambient temperature Storage temperature Relative humidity Location 18 C to 28 C 10 C to 40 C 20 80 non condensing Inside a building 3 2 Performance Data Water flow Water temperature Temperature stability Water reservoir capacity 1 5 1 min 15 20 C 0 5 C 11 3 3 Electric
14. eaid kee atti da 13 DATO STATES Ses ic SKAN e Sa cido 14 6 5 Remotecontrol A aE 14 7 Maintenance Schedull ssccssscssecssssecssssessseesesseesesseseesesesessanseseeenes 16 TIMO UC Ni RR PSP AE EOR 16 T Z AnnualMaintenanceSchedule iaa aoe 16 8 Maintenance Instructions ccscscssssscsssssssseesesseesessesseceseceesansessensesseeanes 17 8 1 Cleaning the Internal Circuit Water Filter c cccccscssessessessessessssssessessessessessessessssussssessessessesssssssussssteseeseeneeneanss 17 8 2 Changing the External Circuit Water Filter Cartridge cccsccscsssssssssssessessesssssssssssssssessessessssnssssesssesssseeneeneenes 18 9 Trouble Sho0tiNd usais 20 10 GE Conformity Notice acacia nara soi 21 Table of Figures Figure 4 1 1 Front View of KMW200B Water Cooler c cccccsscsssssessessessessssesssssssesseesssssssssusssssesseeseeseassansaucstsseeseeseeneeneanss 6 Figure 4 1 2 Back View of the KMW200B Water Cooler c cccccsessssssssssssssssssessessessesssssssusssssessssseesesssansaussuesessseseeneeneanss 6 Figure 4 1 3 Separate front panel of the KMW200B Water Coolet c ccccccsssssssssessssssssstessssessesseesessesssssstsstsssesseneeneanss 7 Figure 5 2 1 Electrical and Water Connections on the Back of the KMW200B onononincincinccnconcnncnnonionoonornrrnornrcnonnonos 10 Figure 5 2 2 Water pipe connection between KMW200B and CCD detectors c c cccsesssssessessessessessstestesseseesees
15. eeneess 11 Figure 5 2 4 Electrical power supply to the KMW200B ccccssssessesssssstssessessessessessssussusstssesseeseessansansausstsseeseeseeneanees 12 Figure 8 1 1 Removing the Water Filter c cccccssssssssessesssscssssessessecsessssssssssucsssessecsesseassaussussucsesseeseessansansaussesseeseeseeneenees 18 Figure 8 2 1 Removing the Water Filter Cartridge c cccccscsssssssssssessessesssssstsssssesssesessesssssssusseesessesseessansanssssesseeseeseeneesens 19 KMW200B Version 15 USER MANUAL Page iv HEALTH AND SAFETY INFORMATION 1 Health and Safety Information 1 1 General In normal operation the system is designed to operate safely All users of the KMW200B water cooler unit should be aware of potential hazards which exist in and around equipment of this type and the ways of avoiding possible injury and equipment damage which may result from inappropriate ways of working A description of such potential hazards and how to avoid them is given in this section This manual adopts the following convention WARNING Indicates a potential hazard which may result in injury or death AN CAUTION Indicates a potential hazard which may result in damage to equipment Warning symbols on the equipment are D Protective conductor terminal Earth ground terminal CAUTION Risk of electric shock CAUTION Refer to accompanying documents gt gt See original manufacturers manuals for furt
16. he back panel of the KMW200B one signals connector and an Ethernet communications connector The power connections are labelled W1 for mains power input and W2 for mains output to drive the CCD detector The Ethernet connector labelled W3 is connected to a LAN card in the PC to provide status information on temperatures flow rate etc The signals connector labelled W4 connects to the separate front panel with the status lamps KMW200B Version 15 USER MANUAL Page 8 HANDLING INSTALLATION STORAGE AND TRANSIT INFORMATION 5 Handling Installation Storage and Transit Information 5 1 Reception and Handling 5 1 1 Delivery The following steps should be carried out on delivery of the system and before unpacking the KMW200B cooler 1 When the system arrives check that there is no visible damage with the delivery driver present If damage has occurred contact the carrier and Agilent Technologies immediately 2 Check that shock watch and tilt indicators fitted to the outside of the packing case have not been activated If the indicators have been activated notify Agilent Technologies immediately 3 Check the number of delivered items against the packing list If any items are missing contact Agilent Technologies within 3 days A WARNING The packing crate is heavy and could cause serious injury and damage to the equipment if not handled correctly Use suitable lifting equipment and procedures Only lift the packing c
17. her safety data on third party equipment supplied with the system A list of these is given in this manual WARNING Do not take risks You have a responsibility to ensure the safe condition and safe operation of equipment 1 2 Electrical Safety In normal use the user is protected from the dangers associated with the voltage current and power levels used by the equipment Only suitably qualified personnel should attempt to disconnect dismantle or modify the equipment WARNING This unit is only to be used when mounted under the electrical rack of an Xcalibur or Gemini machine KMW200B Version 15 USER MANUAL Page 1 HEALTH AND SAFETY INFORMATION 1 2 1 Potential Electrical Hazards The following list is not intended as a complete guide to all the electrical hazards on the system but serves to illustrate the range of potential hazards that exist electric shock electric burn fire of electrical origin electric arcing 1 2 2 Recommended Precautions AN WARNING All of the electrical equipment supplied as part of the system should be provided with a protective ground Do not remove protective grounds as this may give rise to an electrical safety hazard It is vitally important that the system is properly grounded at all times Follow local and national electrical regulations and procedures Do not defeat interlocks remove connectors disconnect equipment open safety covers dismantle o
18. ket W4 on the rear of the KMW200B Connect the Ethernet cable from the system control PC to the socket W3 on the rear of the KMW200B Connect the mains power supply cable from the system mains distributor to the power switch on the separate front panel Connect the mains power supply cable from the power switch on the separate front panel to the rear of the KMW200B on socket W1 Connect the power output socket W2 from the rear of the KMW200B to the CCD detector power supply KMW200B Version 15 USER MANUAL Page 11 HANDLING INSTALLATION STORAGE AND TRANSIT INFORMATION 9 The KMW200B is now ready to use CAUTION When inserting the CCD detector power supply cable into the KMW200B ensure that the connectors are pushed home firmly Mains supply Mains IN CCD detector s power KMW200B 5 supply unit Figure 5 2 4 Electrical power supp 5 3 Storage Before installation commences or when the system is not being used for extended periods store the KMW200B in accordance with the environmental conditions for temperature and humidity stated in the SPECIFICATIONS section of this manual Always store the KMW200B cooler in a secure room KMW200B Version 15 USER MANUAL Page 12 OPERATING INSTRUCTIONS 6 Operating Instructions 6 1 Warnings and Cautions WARNINGS Dangerous voltages are present on the KMW200B s terminals when the power is turned on To prevent personal injury and possible d
19. l specifications Agilent Technologies s liability under this declaration is limited to that set forth in the current Agilent Technologies s Terms and Conditions of Sale Applied Council Directive s 2004 108 EC Electromagnetic Compatibility Directive EMC 2006 95 EC Low Voltage Directive We The Manufacturer Agilent Technologies Poland Sp z 0 0 Ul Rogowska 117B 54 440 Wroctaw Poland declare under our sole responsibility that the following equipment KMW200B Water Cooler Serial Number From CHA 101 07 to which this declaration relates are in conformity with the relevant provisions of the following standard s or other normative document s when installed in conformance with the installation instructions contained in the product documentation EN 55011 Group 1 Class A for radiated emissions EN 55011 Group 1 Class A for mains conducted emissions EN 61326 Electrical equipment for measurement control and laboratory use Pertinent LVD sections of EN 61010 1 amp 2 2001 Safety requirements for electrical equipment for measurement control and laboratory use Part 1 General requirements Part 2 Particular requirements for automatic and semi automatic laboratory equipment Technical Information is maintained at Agilent Technologies Poland Sp z 0 0 Ul Rogowska 117B 54 440 Wroctaw Poland Last two digits of year of CE Marking low Voltage Directive 07 We the undersigned hereby declare that the pr
20. ly depending on the water quality and the length KMW200B operation Tools and Materials Screwdriver Action Personnel Estimated task duration 1 Clean the water filters on the internal and Service technician 30 minutes external water circuits KMW200B USER MANUAL Version 1 5 Page 16 MAINTENANCE INSTRUCTIONS 8 Maintenance Instructions 8 1 Cleaning the Internal Circuit Water Filter Task Time 30 minutes When Once a year dependant on water quality and length of KMW200B operation Tools Screw driver Bucket Small brush Procedure Refer to Figure 8 3 1 1 G2 SL FR ey ot TD Switch off the KMW200B Disconnect the plug from the mains power supply Place a bucket beneath the small filter housing Holding the filter cover unscrew the lower retaining nut Unscrew the filter cover whilst holding the upper retaining nut Remove the filter Clean the filter under clean running water and use the brush as necessary Place the filter back in the filter cover Check the position of the sealing washer is correct inside the upper nut Screw the filter cover into the upper retaining nut Check the position of the sealing washer is correct inside the lower nut Screw on the lower retaining nut Re connect the plug to the mains power supply Switch on the KMW200B Wait for the KMW200B to complete its water conditioning check and then check the water level inside
21. ng operation Water on the external circuit is not cooling the unit correctly too slow flow or too hot water Check pipes are properly connected and valves are open both on the system and at mains water supply point Check mains water temperature is about 15 18C and flows normally Not constant mains water flow in external circuit Water cooler controls flow rate of mains water as and when required No action Air bubbles out of the drain pipe Air is escaping from the closed circuit during automatic water filling Bubbling will stop when filling is completed and should not occur during normal operation Water flows out of the drain pipe Automatic water filling valve is blocked open perhaps due to too high mains water pressure Direct the drain pipe into an open sink Contact Agilent Technologies service Fuses Designators T3 15A 250V Outside on separate front panel T6 3A 250V Boom sio three power supplies KMW200B USER MANUAL Version 1 5 Page 20 CE CONFORMITY NOTICE 10 CE Conformity Notice DECLARATION OF CONFORMITY This Declaration of Conformity is suitable to the European Standard EN 45014 General criteria for supplier s declaration of conformity The basis for the criteria has been found in international documentation particularly in ISO IEC Guide 22 1982 Information on manufacturer s declaration of conformity with standards or other technica
22. nt panel will be illuminated while the KMW200B carries out the water conditioning check 3 After a few seconds check that the CCD detector has been powered lamps lit on the top of the detector and its power supply box 4 The KMW200B is now operational KMW200B Version 15 USER MANUAL Page 13 OPERATING INSTRUCTIONS 5 To monitor the water cooler from the PC connect the Ethernet cable to a LAN card in the PC and set the card s IP address to be fixed and equal to 192 168 254 1 The water cooler has IP address 192 168 254 4 and port number 2500 Telnet can be used to monitor or modify operating parameters 6 4 Error states Short term flow perturbation in the internal circuit will be visible as a continuous flow error indicated on the front panel The unit will switch off after approx 60 seconds of no flow Lack of water inside the internal circuit will start a pulsed filling process which will stop the unit after approx 60 seconds if the unit will be not filled fully No inlet mains water will cause overheating of the water cooling mechanism This state will be signalled as a blinking LED water supply error on the front panel The unit will stop after approx 15 seconds Too hot water inside the unit is signalled by blinking LED temperature error on the front panel The unit will stop after approx 60 seconds Failure in flow level or temperature will be indicated as an error s on the front panel which will remain lit after the
23. oduct s specified above conforms to the listed directive s and standard s Signature Full Name Damian Kucharczyk Position Director XRD Operations Poland Date KMW200B Version 15 USER MANUAL Page 21 Agilent Technologies Inc Printed in Poland September 2010 o Agilent Technologies
24. r modify equipment unless you are qualified and authorised to do so and you are fully conversant with its operation and potential hazards or have total assurance through your local electrical permit to work system that the equipment has been made safe Make sure that the mains supply is fused at an appropriate rating and that it can be isolated locally via a clearly labelled clearly visible and easily accessible isolating switch Isolate the supply before carrying out any maintenance work 1 2 3 First Aid A course in first aid to include methods of artificial respiration is recommended for those whose work involves equipment which may produce a high voltage AN WARNING Do not attempt to administer first aid to someone who may have suffered electric shock until the source of the shock has been isolated Mains voltages are present in the system These can cause serious injury or death Only personnel qualified and experienced to work with such currents and voltages should perform service or maintenance work on this equipment KMW200B Version 15 USER MANUAL Page 2 HEALTH AND SAFETY INFORMATION 1 3 Mechanical Handling Safety WARNING Lifting points are provided for safe handling of components and safe handling practice must be observed to comply with local regulations Check that lifting points are used only for the job intended The system itself and some components are heavy and require careful handling Use
25. the filter housing In order to de aerate the filter unscrew the filter cover momentarily until the water level rises to the top and then re tighten it Check for water leaks If there are any leaks remake the connection KMW200B Version 15 USER MANUAL Page 17 MAINTENANCE INSTRUCTIONS Upper nut containing sealing washer Filter inside housing i Lower nut containing sealing Water inlet hose E F washer Water outlet hose Figure 8 1 1 Removing the Water Filter 8 2 Changing the External Circuit Water Filter Cartridge Task Time 30 minutes When Once a year dependant on water quality and length of KMW200B operation Tools Specialised tool for opening the water filter Bucket It is not advisable to stop the external water supply to the KMW200B while is it operating If it is done then a water supply error will be indicated within about 2 minutes The error state will be reset by turning the unit off and on again However turning off the KMW200B will result in switching off the power to the CCD detector In order to avoid this situation a directional valve and two filters are used Using this valve we can choose which filter is working and which is isolated and could be exchanged Procedure Refer to Figure 8 2 1 1 Change position of the directional valve to isolate the used filter Red arrows show the direction of water flow The KMW200B will now be using the second filter Pl
26. tions Xcalibur is a trademark of Agilent Technologies Limited in some jurisdictions and a registered trademark of Agilent Technologies Limited in other jurisdictions Agilent Technologies acknowledges all trademarks and registrations Copyright 2010 Agilent Technologies Limited All rights reserved No part of this document may be reproduced or distributed in any form or by any means or stored in a database or retrieval system without prior written permission of Agilent Technologies KMW200B Version 15 USER MANUAL Page ii CONTENTS Contents 1 Health and Safety Information 2 1 A A EDT SEERE 1 1 2 Electrical Safety 1 2 1 Potential Electrical Hazards 1 2 2 Recommended Precautions EZS ESCAI A o A 2 1 3 Mechanical Handling Safety cccscsssssssssssssessessessesssssssssessesssssesssssssussussssesseessassanssussssscsesseessansaucsesseeseeaeeseaneass 3 Safe Mechanical race ita 3 1 5 Modifications and Ue ade 3 A A oe 4 AE ARA A A SRS IS E FRE A A O 4 2 2 Howto Use This Mandal iii ad 4 2 3 System DescCriptiON ococononcnnoncnnononnenoneceoenrnrrerrerrererererere rene 4 3 Specifi CANO S adn MENA Md a lio 3 2 Performance Data 3 3 Electrical VIC a 5 3 4 Extertial Water Sup Ns 5 4 Technical Description scsi 6 AAD A ete ee ee ele de eet 6 42 Operating Principles iii dabas 4 2 1 Water Conditioning Check LA essed so 50s tad ect a a Dod a ed sd ta Da Sota eo 4 3 1 Water Supply cc
27. to two separate water circuits an external and an internal The external is connected to the mains water supply and the internal is connected to the device to be cooled In normal operation the two circuits do not exchange water they are connected only by a heat transfer device During start up if the internal circuit has insufficient water even totally dry then it will be automatically filled with water taken from the external circuit After it has been filled this water connection between the circuits is closed The flow rate of mains water in the external circuit is controlled automatically as needed to cool the module Water is pumped around the internal circuit The temperature of water in that circuit is controlled within limits set by internal software and when needed the heat transfer device cools the internal circuit and pumps the heat into the external circuit The temperature of water in the external circuit is also controlled within a second set of limits set automatically and when needed an automatic valve will control the flow of mains water in the external circuit and thus heat will be removed The temperature of water in the external circuit will depend on the temperature of the supplied mains water When the KMW200B is first turned on it will spend a short period analysing the supplied water temperature and then will set control limits accordingly If the mains water supply is cold water lt 15C then only a small flow of mains
28. water will be needed to remove the heat At supply temperatures between 15 and 22C a larger flow of mains supply water will be consumed as necessary to maintain the required temperature in the internal circuit It is recommended that the mains water supply temperature be less than 18C for normal operation AN CAUTION If the mains water supply temperature is above 22C then this is outside of normal specification In this case the maximum flow of mains supply water will be taken and the KMW200B will operate at maximum cooling capacity In this situation the internal circuit temperature will be cooled as much as possible but may not reach the target value To reach normal operation the mains water supply temperature should be reduced to meet the specification 4 2 1 Water Conditioning Check The KMW200B performs a water conditioning check when it is initially switched on This check looks at 1 The level of the water in the internal circuit 2 The flow rate of the water in the internal circuit 3 The temperature of the water in the internal circuit The state of the unit is displayed on four status lights on the separate front panel KMW200B Version 15 USER MANUAL Page 7 TECHNICAL DESCRIPTION 4 3 Interfaces 4 3 1 Water Supply Five water connections are located on the back panel of the KMW200B The connectors are designed for pipes of size 10 x 3 mm 4 3 2 Electrical Connections There are two power connections on t