Agilent 8594Q Option R56 16 256 QAM Retro t Kit Installation Note

image

Contents

1. Floor Mat Figure 1 Example of a Static Safe Workstation 16 and 256 QAM Retrofit Option R56 This kit contains the parts and instructions required to retrofit the 16 and 256 QAM signal processing hardware and Down Loadable Program DLP into the 8594Q Option 190 DVB C QAM Analyzer 16 and 256 QAM Retrofit Option R56 Parts List Parts List Parts included with this kit as supplied Measurement Software DLP card 08594 80048 EEPROM U10 08594 80049 08594 80050 EEPROM U12 5963 6043 Update Information 08594 90096 1 Installation Note and Calibration Guide update This document IMPORTANT Purpose Reason NOTE 16 and 256 QAM Retrofit Option R56 Introduction Introduction Before carrying out the installation of the R56 material ensure you have familiarized yourself with the Safety Considerations and Reliability Considerations at the beginning of this Installation Note The retro fit of 8594Q Option R56 to an 8594Q Option 190 DVB C QAM Ana lyzer To add 16 and 256 QAM constellation sizes to the measurement capability of the 8594Q DVB C QAM Analyzer Before carrying out this procedure ensure you have access to a source of known quality 16 64 and 256 QAM RF signals Cover WARNING Card Cage 16 and 256 QAM Retrofit Option R56 Installation Procedure Installation Procedure The installation of the 8594Q Option R56 Retrofit kit involves fitting replaceme
2. Considerations This is a Safety Class I product provided with a protective earthing ground incorporated in the power cord The mains plug shall only be inserted in a socket outlet provided with a protective earth contact Any interruption of the protective conductor inside or outside the instrument is likely to make the instrument dangerous Intentional disruption is prohibited The opening of covers or removal of parts is likely to expose dangerous voltages Disconnect the instrument from all voltage sources while it is being opened Before switching on this instrument make sure that the line voltage selector switch is set to the voltage of the power supply and the correct fuse is installed DRAFT 3 Protection from Electrostatic Discharge Protection from Electrostatic Discharge Electrostatic discharge ESD can damage or destroy electronic components All work on electronic assemblies should be performed at a static safe workstation Figure 1 shows an example of a static safe workstation using two types of ESD protection e Conductive table mat and wrist strap combination e Conductive floor mat and heel strap combination WARNING These techniques for a static safe workstation should not be used when working on circuitry with a voltage potential greater than 500 volts Building Ground 1 MegOhm Resistor RY Table Mat Wrist Strap fe Building Ground Wrist Strap Cord MegOhm Resistor
3. 8594Q Option R56 16 and 256 QAM Retrofit Kit Installation Note 4 Agilent Technologies Part Number 08594 90096 Printed in USA May 1998 Copyright 1998 Agilent Technologies Inc CAUTION WARNING INSTRUCTION MANUAL WARNING WARNING WARNING WARNING CAUTION Safety Notes Safety Notes The following safety notes are used throughout this Installation Note Familiarize yourself with each of the notesand its meaning before proceeding Caution denotes a hazard It calls attention to a procedure that if not correctly performed or adhered to could result in damage to or destruction of the instrument Do not proceed beyond a caution sign until the indicated conditions are fully understood and met Warning denotes a hazard It calls attention to a procedure that if not correctly performed or adhered to could result in injury or loss of life Do not proceed beyond a warning sign until the indicated conditions are fully understood and met The instruction manual symbol The product is marked with this symbol when it is neccessary for you to refer to the instrucions in the manual These Installation Instructions are for use by qualified personel only To avoid electrical shock do not perform any part of the installation unless you are qualified to do so The power cord is connected to internal capacitors that may remain live for 10 seconds after disconnecting the plug from its power supply General Safety
4. nt of 3 signal processing ROMs to the 8594Q Option 190 DVB C QAM Analyzer and loading new measurement software Carry out the ROM replacement first EEPROM Replacement 1 Ensure all power cables are disconnected from the analyzer 2 Protect the front face of the analyzer and place it face down on a smooth work surface 3 Loosen the 4 Allen headed fixing bolts on the rear of the analyzer 4 Loosen and remove the 4 screws that fix the analyzer cover to the rear panel The spectrum analyzer contains potentially hazardous voltages Ensure you are familiar with the general safety considerations at the front of this Installation Note 5 Remove the analyzer cover 6 Remove the Ribbon Cable connecting the Digital Signal Processor card A190A3 to the Digital Demodulation card A190A2 DSP card A190A3 Ribbon cables Front of Analyzer 7 Remove the Digital Signal Processor card from the card cage 16 and 256 QAM Retrofit Option R56 Installation Procedure 8 Remove the 3 ROMs U10 U11 and U12 and replace each with the respective device supplied with this kit Ensure each device is correctly oriented and located r i U 10 a U 12 _ 9 Refit the Digital Signal Processing card into the card cage and refit the Ribbon Cable connecting to the Digital Demod card Cover 10 Refit the analyzer cover Take care to avoid snagging any loose cables 11 Fix the 4 scre
5. ws and 4 bolts to the rear of the analyzer The assembly of the analyzer is now complete Measurement Software Replacement and Analyzer Verification The next procedure details the installation of the new measurement software DLP and verifies the correct operation of the analyzer CAUTION Do not connect AC power until you have verified that the line voltage is correct the proper fuse is installed and the line voltage selector switch is properly positioned Damage to the analyzer could result Delete old DLP 1 Press to power up the analyzer 2 Erase the currently installed measurement DLP by pressing CONFIG More 1 of 3 Dispose User Mem ERASE DLP MEM Press ERASE DLP MEM again to delete the DLP 3 Press PRESET to initialize the analyzer Install new DLP Verification 16 and 256 QAM Retrofit Option R56 Installation Procedure 4 Insert the DLP card into the analyzer card reader 5 Load the new measurement DLP by pressing RECALL INTERNAL CARD to underline CARD Cataleg Card catalog ATI Ensure the file ADCATV_D is highlighted and press LOAD FILE to load the DLP 6 Connect the analyzer to the QAM source and confirm the constellations lock at 16 64 and 256 QAM On successful completion of Verification the analyzer is ready for use The supplied Update Information 5963 6043 provides brief details of the new features 10